Installation Completion and System Verification 3
Step 4. Set up methods and worklists
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 61
The worklist is set up to:
• Do one injection of the solvent blank with the scan method to collect
background ion data.
• Use the MRM method for the remaining runs to do:
• one injection of the empty vial
• five injections of the solvent blank
• five injections of the checkout sample
Tabl e 3 Checkout methods and worklists
Model Mode Positive Ion Mode
Installation Checkout
Negative Ion Mode
Installation Checkout
6495 Scan Method 6495 Pos MS2 Scan.m 6495 Neg MS2 Scan.m
MRM Method 6495 Pos MRM Checkout.m 6495 Neg MRM Checkout.m
Worklist 6495 Pos MRM Checkout.wkl 6495 Neg MRM Checkout.wkl
6490 Scan Method 6490 ESI_AJT Pos MS2 Scan Grad.m 6490 ESI_AJT Neg MS2 Scan.m
MRM Method 6490 ESI_AJT Pos MRM Checkout Grad.m 6490 ESI_AJT Neg MRM Checkout.m
Worklist 6490 ESI_AJT Pos MRM Checkout Grad.wkl 6490 ESI_AJT Neg MRM Checkout.wkl
6470 Scan Method 6470 Pos MS2scan.m 6470 Neg MS2scan.m
MRM Method 6470 Pos MRM Checkout.m 6470 Neg MRM Checkout.m
Worklist 6470 Pos MRM Checkout.wkl 6470 Neg MRM Checkout.wkl
6460
(AJS)
Scan Method 6460 ESI_AJT Pos MS2 Scan Grad.m 6460 ESI_AJT Neg MS2 Scan.m
MRM Method 6460 ESI_AJT Pos MRM Checkout Grad.m 6460 ESI_AJT Neg MRM Checkout.m
Worklist 6460 ESI_AJT Pos MRM Checkout Grad.wkl 6460 ESI_AJT Neg MRM Checkout.wkl
6460
(ESI)
Scan Method 6460 ESI Pos MS2 Scan Grad.m 6460 ESI Neg MS2 Scan.m
MRM Method 6460 ESI Pos MRM Checkout Grad.m 6460 ESI Neg MRM Checkout.m
Worklist 6460 ESI Pos MRM Checkout Grad.wkl 6460 ESI Neg MRM Checkout.wkl
6420/
6430
(ESI)
Scan Method 6400 ESI Pos MS2 Scan Grad.m 6400 ESI Neg MS2 Scan.m
MRM Method 6400 ESI Pos MRM Checkout Grad.m 6400 ESI Neg MRM Checkout.m
Worklist 6400 ESI Pos MRM Checkout Grad.wkl 6400 ESI Neg MRM Checkout.wkl