3 Installation Completion and System Verification
Step 5. Verify the Triple Quadrupole LC/MS sensitivity
68 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
Figure 35 Example of a Negative mode signal-to-noise report.
6 Change solvents (to A = water w/0.1% formic acid and B = acetonitrile for
Positive Ion Polarity).
7 Clean the injectors. See step 12 on page 52.
8 Equilibrate the column for 1 hour at 30:70 A:B.