5 Other Setup Tasks
To reconfigure the instrument
90 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
6 Click OK.
7 Start the MassHunter Workstation Acquisition program and confirm that
only the MS instrument is available.
8 Close the Data Acquisition program.
9 Stop MassHunter processes.
10 Open Instrument Configuration again.
11 Click the Triple Quadrupole LC/MS and mark the Agilent LC (1100, 1120,
1200, 1220, 1260, 1290)/CE check box.
Figure 38 Instrument Configuration dialog box
12 Click OK.
The registry has now been cleared and configured for any new LC module(s).
Additional configuration steps may be required to set up LC devices.