Hardware Symptoms 2
Troubleshooting 13
Vial Mishandled by ALS
Refer to the 7683B Automatic Liquid Sampler Installation,
Operation and Maintenance manual for additional information.
When you find a mishandled sample vial, do the following:
Check for folds or wrinkles in the crimp cap, especially near
the neck of the sample vial.
Use Agilent recommended sample vials.
Check sample labels (if applicable).
Check that they are the correct size.
Verify that the labels do not interfere with the gripper.
Check that the tray quadrants are clean and snapped into the
tray base.