96 Troubleshooting
8 Troubleshooting Tasks
To Measure FID Leakage Current
1 Load the analytical method.
• Make sure flows are acceptable for ignition.
• Heat the detector to operating temperature or to 300 °C.
2 Turn off the FID flame.
3 Verify that the FID electrometer is on.
4 Press [Front Detector] or [Back Detector], then scroll to Output.
5 Verify that the output is stable and < 1.0 pA.
If the output is unstable or > 1.0 pA, turn off the GC and
check for proper assembly of the upper FID parts and for
contamination. If the contamination is confined to the
detector, bakeout the FID.
6 Turn on the flame.