Chromatographic Symptoms 3
Troubleshooting 41
FPD Output Too High or Too Low
Verify that the correct filter is being used. Do not use a
phosphorus filter with sulfur-optimized flows or a sulfur
filter with phosphorus-optimized flows.
Check the position of the column as installed in the detector.
Check the gas purity.