Chromatographic Symptoms 3
Troubleshooting 49
NPD Low Selectivity
Verify that the hydrogen flow is correct (< 3 mL/min).
Inspect the bead; it may be defective or expended.
Verify correct bead voltage.
Replace the collector and insulators.