70 Troubleshooting
5 Shutdown Symptoms
Column Shutdown
If a carrier gas (which can include an auxiliary flow module or
pneumatics control module) shuts down, the following happens:
The oven turns off to avoid column damage.
The oven flaps in the back of the oven open halfway.
The oven temperature setpoint flashes Off.
All flows for the column turn off. When viewed, their
parameters flash Off. For example, the septum purge and
column flows for a split/splitless inlet would turn off.
All other heaters turn off. When viewed, their temperature
parameters flash Off.
Attempts to turn on a shut-down zone fail with an error
message.
The instrument beeps.
To recover from this state.
1 Fix the cause of the shutdown.
Check for a broken column.
Replace the inlet septum.
Replace the inlet O-ring.
Check the supply pressure.
Check for leaks.
2 Press the key for the device that initiated the shutdown.
Scroll to the pneumatic parameter that is flashing Off, then
press [On] or [Off].
For example, if the front inlet ran out of carrier gas, press
[Front Inlet], scroll to the pressure or flow parameter, then
press [On].

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890N and is the answer not in the manual?

Agilent Technologies 6890N Specifications

General IconGeneral
BrandAgilent Technologies
Model6890N
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals