Chromatographic Symptoms 3
Troubleshooting 27
Baseline Rise During Oven Temperature Program
Inspect the column for bleed.
Check for leaks/oxygen in carrier gas supply.
Check gas supply oxygen trap indicator or date.
Make solvent blank runs to evaluate baseline without sample.
Make “no injection” blank runs (remove the syringe from the
injector and start a run) to evaluate baseline without solvent.
Check for contamination. (See Contamination or Carryover.)
Consider the effect of column film thickness on bleed.
Check for leaks at the column fittings. (See “Checking for
Leaks” .)
Prepare and use a column compensation profile.