Chromatographic Symptoms 3
Troubleshooting 31
Sample Decomposition in Inlet/Missing Peaks
Lower the inlet temperature.
Check for air or water in the carrier gas; verify gas purity and
functionality of traps.
Verify that the liner is appropriate for the sample being run.
Perform complete inlet maintenance: Replace all consumable
parts and bake out the inlet.
Install a deactivated liner (SS, PP, and PTV inlets).
Check for leaks at the septum, liner, and column fittings. (See
“Checking for Leaks” .)
Install an Agilent Direct Connect liner.
Use a pulsed pressure method for quicker sample transfer to
column.
Bake out the inlet. See the following:
To Bakeout Contaminants from the Split/Splitless Inlet
To Bakeout Contaminants from the Purged Packed Inlet
To Bakeout Contaminants from the COC Inlet
To Bakeout Contaminants from the PTV Inlet
To Bakeout Contaminants from the VI Inlet