GC Not Ready Symptoms 4
Troubleshooting 55
Flow Never Becomes Ready
If the gas flow never becomes ready, check for the following:
Check the supply gas for sufficient delivery pressure.
Check the configured gas type. The configured gas type must
match the actual gas plumbed to the GC.
Check for leaks in the gas delivery plumbing and the GC. (See
“Checking for Leaks” .)