EasyManua.ls Logo

Agilent Technologies 6890N - FID Flame Goes out During a Run and Attempts to Reignite

Agilent Technologies 6890N
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38 Troubleshooting
3 Chromatographic Symptoms
FID Flame Goes Out During a Run and Attempts to Reignite
The following is an example chromatogram showing a flameout
from a large solvent peak.
After a flameout, the GC will try to ignite the flame three times.
If the FID flame goes out during a run, do the following:
See if an aromatic peak or water extinguished the flame.
Check for a plugged jet.
Verify that the gas flow settings are correct. Verify that Lit
offset is set appropriately.
If the FID flame attempts to reignite but is already lit, do the
following:
Verify that the FID Lit offset setting is appropriate for the run
(typically
< 2.0 pA).
Check to see if an aromatic peak or water extinguished the
flame.
Check for a partially plugged jet. Measure actual hydrogen,
air, and makeup flows at the detector. (See “To Measure a
Detector Flow” .)
Check for leaks at the detector column fitting. (See “Checking
for Leaks .)
9
8
7
6
5
4
3
2
1
23
4
5min
pA

Table of Contents

Related product manuals