Chromatographic Symptoms 3
Troubleshooting 25
Larger Peaks Than Expected
Check each configured column’s dimensions against the
actual column dimensions. (See “Configurable Items to
Always Keep Current” .)
Check the autosampler injection volume.
Check the vial caps.
Check configured syringe size. Some syringe sizes are
specified at half-capacity. If the maximum syringe volume is
marked at half-height on the barrel, not at the top of the
barrel, enter twice the labeled volume when configuring the
syringe size.