EasyManuals Logo

Agilent Technologies 6890N User Manual

Agilent Technologies 6890N
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
84 Troubleshooting
8 Troubleshooting Tasks
To Measure a Column Flow
Measuring FID, TCD, uECD, and FPD column flow
The following procedure can be used to measure column flow
with an FID, TCD, uECD, and FPD.
1 Gather the following:
Appropriate flowmeter adapter tube (can be found in the
GC ship kit)
Electronic flowmeter calibrated for the gas and flow rates
of concern
2 Turn off the detector.
3 Turn off the detector flows.
4 Connect the appropriate adapter to the detector exhaust.
A 1/8-in rubber adapter tube attaches directly to a uECD or
TCD exhaust vent.
A separate adapter (19301-60660) is supplied for the FID.
Insert the adapter into the detector exhaust vent as far as
possible. You will feel resistance as the adapter O-ring is
forced into the detector exhaust vent. Twist and push the
adapter during insertion to ensure a good seal.
For the FPD, remove the vent tube assembly and use a 1/4-in
tube adapter between the detector exhaust and the
flowmeter tubing.
WARNING
Hydrogen (H
2
) is flammable and is an explosion hazard when
mixed with air in an enclosed space (for example, a flow meter).
Purge flowmeters with inert gas as needed. Always measure
gases individually. Always turn off detectors to prevent
flame/bead autoignition.
WARNING
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890N and is the answer not in the manual?

Agilent Technologies 6890N Specifications

General IconGeneral
BrandAgilent Technologies
Model6890N
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals