GC Not Ready Symptoms 4
Troubleshooting 67
NPD Adjust Offset Process Fails
Inspect the jet to see if it is clogged.
Measure the actual detector flows. (See To Measure a
Detector Flow.)
Check the condition of the bead. Replace if necessary.
Verify that the flow settings are correct.
Thoroughly leak check the whole system, especially the
detector column fitting. (See Checking for Leaks.)
Set the equilibration time to 0.0.