30 Troubleshooting
3 Chromatographic Symptoms
Peak Boiling Point or Molecular Weight Discrimination Poor
If you have trouble with peak boiling point or molecular weight
discrimination (inlet discrimination), do the following:
Check the inlet for contamination. Clean and change the
liner if necessary. Replace all inlet consumable parts. See the
Maintenance manual.
Adjust the inlet temperature.
Run standards against a known method to determine
expected performance.
For any inlet operating in split mode with any detector
Check liner type.
Increase the inlet temperature and verify that the insulation
cup is installed and contains insulation.
Check column cut and installation into the inlet. See the
topic for the SS, PTV, and VI.
For any inlet operating in splitless mode with any detector
Check the inlet for leaks. (See “Checking for Leaks” .)
Check liner type.
Verify that the oven starting temperature is less than the
solvent boiling point.
Check column cut and installation into the inlet See the topic
for the SS, PTV, and VI.
Check that the solvent vapor volume does not exceed the
liner capacity.
Check for appropriate purge delay time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890N and is the answer not in the manual?

Agilent Technologies 6890N Specifications

General IconGeneral
BrandAgilent Technologies
Model6890N
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals