84 708-DS / 709-DS Operator’s Manual
4 Operating the 708-DS / 709-DS
Menu Screen - Instrument Settings Screen 2
Table 7 Menu Screen - Instrument Settings Screen 2
Option Function
Temperature
Tolerances
Allowable temperature variation from the set point. When alarms are
enabled in Instrument Settings, deviations greater than this will cause an
alarm but not stop the method.
Speed Tolerance Allowable spindle speed variation from the set point. When alarms are
enabled in Instrument Settings, deviations greater than this cause an
alarm but do not stop the method.
Vessels This parameter sets the current installed vessel size (2000 mL, 1000 mL,
200 mL, or 100 mL). This is used to set user prompts in other places
throughout the operation of the instrument.
Number Active Sets the number of vessels used in a test. For example, a 6-vessel
configuration may only use the first four vessels if this parameter is set
to 4. The remainder of the vessels are ignored until this parameter is
changed.
Enable Alarms If checked, all alarms are activated. If not checked, all alarms are
suppressed on the user interface and on a printout if a printer is installed.
Enable DDM Enables the Dosage Delivery Modules. This setting overrides the
method setting. If DDM is not enabled, the DDM Increment box is
grayed out in Method Editor.