Faults
27
No Light Is On
Probable causes
The line voltage to the GC is off.
The injector cable or connection to the GC is bad.
Your GC requires service.
Suggested actions
1. Verify the injector is properly connected to the GC.
2. Check the power source for your GC.
3. Obtain Agilent service.
The Fault Light Is On
Probable cause
The injector door is open.
Suggested actions
1. Ensure that the injector door is closed.
2. If the fault light stays on, obtain Agilent service.
The Fault And The Run Lights Are On
Probable causes
The injector is mounted incorrectly on the mounting post.
Incorrect mounting post.
Suggested actions
1. Ensure that the injector is mounted properly. For more information,
see page 4.
2. Ensure the correct mounting post is installed. See page 4.
3. If the fault lights stay on, obtain Agilent service.