Installation Guide
Error Messages
35
No Bottle In Gripper
Probable causes
The sample vial was not found by the gripper.
The gripper could not grasp the vial.
The vial was dropped during transfer to or from the turret.
The sensor in the gripper is defective.
Suggested actions
1. Make sure that the sample vials are in the locations specified by the
sequence.
2. Make sure that the sample vials meet recommended specification. See
your Sampling Techniques Handbook.
3. If you are using adhesive labels verify that the labels are properly
installed. See your Operation Guide.
4. If the error recurs frequently, obtain Agilent service.
Tray Not Present
Probable causes
There is a board failure in the tray or GC.
The tray cable is bad or not connected between the GC and tray.
There is a cable failure in the GC.
Suggested actions
1. Make sure that the tray cable connection is secure.
2. Replace the tray cable.
3. If the error remains, obtain Agilent service.

Other manuals for Agilent Technologies 7683

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7683 and is the answer not in the manual?

Agilent Technologies 7683 Specifications

General IconGeneral
BrandAgilent Technologies
Model7683
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals