Installation Guide
Error Messages
37
No Injector
Probable causes
The cabling connection to the GC became loose during a run.
An injector board or GC board failed during a run.
Suggested actions
1. Make sure that the connection to the GC is secure.
2. If the error remains, obtain Agilent service.
Prerun > 10 Min
The GC is Not Ready. Check for Not Ready and other GC messages to determine
the cause.
Sampler Error
The sampler could not function for an undocumented reason. Record the code
number shown in the G2629A Control Module error message. If the problem
persists, obtain Agilent service and report the error number.