Installation Guide
Error Messages
36
Tray Offline
Probable causes
• There is a board failure in the tray or GC.
• The tray cable is bad or not connected.
• There is a cable failure in the GC.
Suggested actions
1. Make sure that the tray to GC cable connection is secure.
2. Replace the tray cable.
3. If the error remains, obtain Agilent service.
Autoinject Aborted
The autoinject sequence on the 6850 was aborted. The other error messages
shown on the GC display provide more information on what caused the sequence
to abort.
Invalid Sequence
Probable causes
• The sequence is set up for the wrong injection device.
• Hardware required by the sequence is not installed and configured.
• The GC configuration was changed during sequence execution.
• The injector cable is bad or not connected properly.
Suggested actions
1. Make sure that the connection to the GC is secure.
2. Verify the sequence parameters against the GC configuration.
3. If the error remains, obtain Agilent service.