Installation Guide
Faults
28
Align Mode Light Is On
Probable causes
The turret is not properly installed.
The type of turret was changed while the power was on.
The system was not initialized.
There is an injector memory error.
Suggested action
1. Verify the turret is properly installed. See Removing The Turret on
page 16.
2. Perform the alignment procedure to initialize the system. See Turret
Alignment on page 17.
All Lights Are On
Probable causes
The cables connections are loose.
There is a board failure.
There is a firmware revision conflict.
Suggested action
1. Check all cable connections.
2. Turn the instrument off, then on again.
3. If the lights remain on, obtain Agilent service.