Installation Guide
Adapting For Cool On-column Injection
10
Adapting For Cool On-column Injection
The 7683 Automatic Liquid Sampler can be adapted to inject samples directly
onto 250-
µm, 320-µm, and 530-µm columns in the 6890 with cool-on-column inlet.
If you wish to adapt the injector and GC for cool on-column use, follow these
steps:
1. Select the on-column syringe needed for the column size. See your Agilent
catalog for consumables and supplies, and Volume 2 of your GC Operating
Manual for a list of parts.
2. Prepare the GC inlet. (See Volume 2 of your GC Operating Manual for
instructions).
Check your needle to column size.
Verify that the insert matches the needle size.
If necessary, replace the septum.
3. Verify that the correct needle support assembly is installed in the injector.
Replace if necessary. See Replacing The Needle Support Assembly on
page 19.
For a list of consumables needed to perform these injections, see Volume 2 of
your GC Operating Manual.

Other manuals for Agilent Technologies 7683

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7683 and is the answer not in the manual?

Agilent Technologies 7683 Specifications

General IconGeneral
BrandAgilent Technologies
Model7683
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals