Error Messages
32
Front (Or Back) Plunger Error
Probable causes
The syringe plunger is sticking, or not securely connected to the plunger
carrier.
The plunger solenoid is binding.
The plunger carrier encoder is inoperable.
Suggested actions
1. Remove the syringe, and check it for plunger stickiness or binding.
Replace the syringe if necessary. For more information, see Inspecting
a syringe in the Operation Guide.
2. Check the viscosity of the sample against the viscosity parameter. Reset
the viscosity parameter if necessary.
3. Restart the sequence.
4. If the error occurs again, obtain Agilent service.
Front (Or Back) Syringe Error
Probable causes
The syringe carriage motor is defective.
The syringe is not currently installed or is an incorrect type.
The syringe carriage sensor is inoperable.
Suggested actions
1. Ensure the syringe is installed correctly. For more information, see your
Operation Guide.
2. Ensure the syringe meets specifications.
3. If the syringe needle is bent, see Correcting Syringe Problems on
page 24.

Other manuals for Agilent Technologies 7683

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7683 and is the answer not in the manual?

Agilent Technologies 7683 Specifications

General IconGeneral
BrandAgilent Technologies
Model7683
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals