Installation Guide
Error Messages
34
2. Check your method to make sure it uses the appropriate injector
location.
3. If the error remains, obtain Agilent service.
Injector Offline
Probable causes
There is a board failure in the injector or GC.
The injector cable is bad or not connected.
There is a cable failure in the GC.
Suggested actions
1. Make sure that the injector to GC cable connection is secure.
2. If the error remains, obtain Agilent service.
No Bar Code Reader
Probable causes
Bar code reader cable not securely connected.
Bar code reader is defective.
Tray is defective.
Suggested actions
1. Ensure the bar code reader cable connection is secure.
2. If the problem continues, obtain Agilent service.