Installation Guide
Error Messages
30
Bottle In Gripper
Probable cause
The sample vial was not delivered properly and stayed in the tray gripper.
Suggested actions
1. Remove the vial and return it to its position in the tray.
2. Ensure that the tray quadrants are snapped into place.
3. Ensure that the injector is plugged into the correct connector on the
back of the 6890 GC and is configured properly. See Configuring Your
Gc (6890) on page 12.
4. Check the “deliver to” location for the vial and verify that the location
is empty and free from obstructions.
5. Make sure that the injector is upright/vertical on the GC.
6. Restart the sequence.
7. If the error occurs again, obtain Agilent service.
Front (Or Back) Door Open Or Injector Not Mounted
See Faults on page 26.
Front (Or Back) Injector Com Error
Probable cause
There is a communications error between the injector and the GC.
Suggested action
Obtain Agilent service.

Other manuals for Agilent Technologies 7683

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7683 and is the answer not in the manual?

Agilent Technologies 7683 Specifications

General IconGeneral
BrandAgilent Technologies
Model7683
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals