12 Advanced User Guide
2 Programming
FINAL DRAFT - 8 AUG 2016
Run Time Programming
Run time programming during a method allows certain
setpoints to change automatically during a run as a function of
the chromatographic run time. Thus an event that is
programmed to occur at 2 minutes will occur 2 minutes after
every injection.
Its uses include:
Controlling column switching or other valves
Changing analog signal definition, zero, or range
Controlling an auxiliary pressure channel
Changing polarity of a thermal conductivity detector (TCD)
Turning the hydrogen flow to a nitrogen-phosphorus
detector (NPD) on or off
Switching digital signal output (requires an Agilent data
system)
Pausing (“freezing”) and resuming digital signal output
(requires an Agilent data system)
The changes are entered into a run table that specifies the
setpoint to be changed, the time for the change, and the new
value. At the end of the chromatographic run, most setpoints
changed by a run time table are returned to their original
values.
Valves can be run time programmed but are not restored to
their starting position at the end of the run. You must program
the reset operation in the run table if this action is desired. See
“From the run or clock time tables” on page 115.
Using run time events
The [Run Table] key is used to program the following timed
events.
Valves (1-2)
Analog signal definition, zero, and range
TCD negative polarity (on/off)
Detector gas flow (on/off), including NPD H
2
fuel gas

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals