Inlets 3
Advanced User Guide 39
About the Purged Packed Column Inlet
This inlet is used with packed columns when high-efficiency
separations are not required. It can also be used with wide-bore
capillary columns, if flows greater than 10 mL/min are
acceptable.
The inlet is flow controlled, regardless or whether or not the
column is capillary and defined, or packed.
Valve
Frit
EPC Module
Inlet Weldment
FS = Flow Sensor
PS = Pressure Sensor
FS
Carrier Supply
80 PSI
Septum Purge
Column
PS
Frit