Programming 2
Advanced User Guide 13
FINAL DRAFT - 8 AUG 2016
Programming run time events
1 Press [Run Table].
2 Press [Mode/Type] to see the available run time events.
3 Scroll to the event to be programmed. Press [Enter].
4 Enter values for the Time: and the other parameter. Press 
[Enter] after each entry.
5 Press [Mode/Type] to add another event. Press [Status] to 
terminate entries.
The run table
The programmed events are arranged in order of execution time 
in the Run Table. This is a brief example:
RUN TABLE  (1 of 3) Event 1 rotates a valve. 
Time: 0.10 
Valve #2 On 
RUN TABLE  (2 of 3) Event 2 adjusts the signal 
Time: 3 range. 
Analog signal 2 range 2 
RUN TABLE  (3 of 3) Event 3 resets Valve #2 to its 
Time: 4.20 original position in preparation 
Valve #2 Off for another run. Valves do not  
reset automatically.
Adding events to the run table
1 To add new events to the run table, press [Mode/Type] while 
on the Time: line of any entry.
2 Select the event type.
3 Set appropriate Time: and other parameters. Some require 
numbers; others require [On/Yes] or [Off/No].
4 Repeat until all entries are added. Events are automatically 
placed in order by execution time.