Splitless liner
The liner volume must contain the solvent vapor. The liner
should be deactivated to minimize sample breakdown during
the purge delay. Solvent vapor volume can be reduced by using
Pulsed Splitless mode. Use the “Vapor Volume Calculator“ to
determine vapor volume requirements.
Vapor volume < 300 µL Use 2 mm liner (250 µL volume),
5181-8818 or similar.
Vapor volume 225 – 300 µL Consider pulsed splitless mode to
reduce vapor volume.
Vapor volume > 300 µL Use 4 mm liner, 5062-3587 or similar.
Vapor volume > 800 µL Consider pulsed splitless mode to
reduce vapor volume.
For thermally labile or reactive samples, use G1544-80700 (open
top) or G1544-80730 (top taper) liners.
NOTE
Pulsed split and pulsed splitless modes are not supported on EPR
(electronic pneumatics regulation) equipped GCs.
Table 5 Splitless mode liners
Liner Description Volume Mode Deactivated Part Number
Single Taper Glass
Wool
900 uL Splitless Yes 5062-3587
Single Taper 900 uL Splitless Yes 5181-3316
Dual Taper 800 uL Splitless Yes 5181-3315
2 mm Quartz 250 uL Splitless No 18740-80220
2 mm Quartz 250 uL Splitless Yes 5181-8818
1.5 mm 140 uL Direct
Inject, Purge
and Trap,
Headspace
No 18740-80200
Single Taper Glass
Wool
900 uL Splitless Yes 5062-3587