60 Advanced User Guide
4 Columns and Oven
Position 2 Carrier gas flows through the column toward the
inlet, removing components on the column through the inlet
vent line.
The Run Table contains commands to perform these actions:
After the last peak of interest appears, switch the valve to
Position 2. Higher boiling peaks are discarded through the
inlet vent.
At the same time, turn data acquisition off.
At the end of the backflush period (determined
experimentally), switch the valve to Position 1 and turn data
acquisition on. The system is now ready for the next run.