EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8562EC

Agilent Technologies 8562EC User Manual

Agilent Technologies 8562EC
716 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #315 background imageLoading...
Page #315 background image
Chapter 5 315
Programming
Data Transfer to Computer
Transmission Sequence of Data on GPIB
Table 5-2 on page 315 shows a GPIB transmission sequence for each
format mode. Each one transmits the +10 dBm amplitude level of a
one-element trace with the amplitude equal to the reference level. In
each case, the HP 9000 Series 200 or 300 computer must be instructed
how to interpret the data received correctly. The parenthetical numbers
in the table are decimal values representing binary 8-bit numbers.
Table 5-2 GPIB Transmission Sequence for Different Trace-Data Formats
(TDF)
Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6
TDF P
1 0 . 0 0 (10-EOI)
TDF M
6 0 0 (10-EOI)
TDF A
# A (0) (2) (2) (88-EOI)
TDF I
# I (2) (88-EOI)
TDF B
(2) (88-EOI)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8562EC and is the answer not in the manual?

Agilent Technologies 8562EC Specifications

General IconGeneral
BrandAgilent Technologies
Model8562EC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals