EasyManuals Logo

Agilent Technologies 89410A Service Guide

Agilent Technologies 89410A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
2-8
Adjusting the Analyzer
To adjust oven shutdown
2 Set up the spectrum analyzer.
a Set the spectrum analyzer as follows:
b Connect the spectrum analyzer to the Agilent 89410A’s OVEN
REF OUT connector using the BNC cable.
3 Check that the 10 MHz signal is
-70 dBm.
a Set the power switch to on ( l ).
b Check that the amplitude of the 10 MHz signal is
-70 dBm (oven
is cold).
4 Adjust A85 R2.
a Change the spectrum analyzer’s reference level to 10 dBm.
b Turn A85 R2 counterclockwise until the 10 MHz signal’s
amplitude is > 0 dBm.
c Turn A85 R2 clockwise until the signal’s amplitude just returns to
approximately -70 dBm, then turn R2 an additional 10 degrees
clockwise.
Center Frequency
Frequency Span
Reference Level
10 MHz
1MHz
60 dBm

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals