Replacing Assemblies
To place the A82 in the test position
5
Unplug the 3 pin connector from the A82 Flat Panel
Interface assembly.
6
Using a T-10 torx driver, remove the six screws from the A82
sheild. Remove the A82 shield.
7
Reconnect the 3 pin connector to the A82 Flat Panel
Interface assembly.
8
Route W4, W6, and W17 to the top and reconnect to the
front panel assembly. W4 will need to be unplugged from
A90P10 in order to route it under the air dam.