EasyManuals Logo

Agilent Technologies 89410A Service Guide

Agilent Technologies 89410A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
1-78
Troubleshooting the Analyzer
To troubleshoot two channel analyzer failures
4 Check the analog source to input paths.
a Press the following keys:
[
Preset]
[
Source]
[source
on]
[
level]
0
[
dBm]
[
System Utility]
[more cal setup]
[
auto zero cal off]
[
Return]
[
more]
[
diagnostics]
[
service functions]
[
special test modes]
[
input]
[
channel 1 source]
b Using a BNC(m)-to-SMB(f) cable, connect the oscilloscope to
channel 1 A10 J2.
The oscilloscope should display a 130 ±13 mVp-p sine wave.
c Reconnect the cable from channel 1 A10 J2 to A21 J4.
d Press the following keys:
[channel 1 front BNC]
[
channel 2 source]
e Using a BNC(m)-to-SMB(f) cable, connect the oscilloscope to
channel 2 A10 J2.
The oscilloscope should display a 130 ±13 mVp-p sine wave.
f Reconnect the cable from channel 2 A10 J2 to A21 J4.
g If both signals are correct, go to Step 5.
h Place the analyzer on the side closest to the display.
i Remove the bottom cover.
j Set the oscilloscope as follows:
CH1 V/div 100 mV/div
Input Impedance 1 M
CH1 Coupling dc

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals