Harmonic Distortion
Measurement Lower Limit Upper Limit
(dBc)
Measured Value
(dBc)
Pass/Fail
9.75 MHz 2nd, Ch 1
−75
9.75 MHz 3rd, Ch 1
−75
9.75 MHz 2nd, Ch 2 †
−75
9.75 MHz 3rd, Ch 2 †
−75
† Option AY7 only
DC Offset
Measurement Lower Limit Upper Limit
(dBfs)
Measured Value
(dBfs)
Pass/Fail
−30dBm,Ch1,FilterIn
−25
−20dBm,Ch1,FilterIn
−25
−10dBm,Ch1,FilterIn
−25
0 dBm, Ch 1, Filter In
−25
+10 dBm, Ch 1, Filter In
−25
+20 dBm, Ch 1, Filter In
−25
−30dBm,Ch2,FilterIn†
−25
−20dBm,Ch2,FilterIn†
−25
−10dBm,Ch2,FilterIn†
−25
0 dBm, Ch 2, Filter In †
−25
+10 dBm, Ch 2, Filter In †
−25
+20 dBm, Ch 2, Filter In †
−25
† Option AY7 only
Serial Number:______________________Report Number:____________________
Agilent 89410A Verifying Specifications
Operation Verification Test Record
5of8