1200 Series DAD and MWD User Manual 121
Maintenance 8
5 Use a small flat screw driver to carefully lift off the I.D. tag.
Shown is the default orientation. See Note on page 114.
6 Carefully insert the I.D. tag into the new heat exchanger.
Shown is the default orientation. See Note on page 114.
7 Fix the new heat exchanger to the clamp unit and the heat
exchanger to the cell body.
8 Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4-mm wrench.