142 1200 Series DAD and MWD User Manual
8Maintenance
Test Failed (low ratio value)
Probable Causes
Absorbing solvent or air bubble in flow cell.
Dirty or contaminated flow cell.
Suggested Actions
Ensure the flow cell is filled with water, and free from air bubbles.
Exchange the flow cell windows.