1200 Series DAD and MWD User Manual 47
Using the Detector 4
Conditions
A single injection of the isocratic test standard is made under the conditions
given in Table 7:
Tab le 7 Conditions
Flow 1.5 ml/minute
Stoptime 8 minutes
Solvent 100% (30% water/70% Acetonitrile)
Temperature Ambient
Wavelength sample 254 nm (4 nm bandwidth)
reference 360 nm (100 nm bandwidth)
Injection Volume 1 µl