Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 111
Optimizing the Detector
5
Example: Optimization for Multiple Compounds
4 Using the settings and the timetable (from previous page), do a second run
for the evaluation of the optimal excitation wavelength. See Figure 40 on
page 111.
Table 15
Peak # Time Emission Wavelength
15.3min330 nm
27.2min330 nm
37.6min310 nm
48.6min360 nm
5 10.6 min 445 nm
6 11.23 min 385 nm
DO NOT select
additional excitation
wavelengths (B, C,
D).
Doing so will
increase the scan
time and will lower
the performance.
Figure 40 Detector Settings for Excitation Scan