Agilent InfinityLab LC Series Vialsamplers User Manual 102
3Using the Module
Using the Optional Integrated Column Compartment
Column Assignment
Clicking on Column Assignment in the control interface opens a dialog box, where
the column information can be edited:
• Plumbing: The user can specify whether the column is located in the Upper or
Lower position of the ICC heat exchanger, or indicate that no column is
present in the ICC at all (None).
• Visualization: This section gives a visual representation of the column
configuration in accordance with the information in the Plumbing section.
• Column Tag Information: This section holds information on the physical
properties of the columns installed in the ICC and their usage (see
“Column
Tag Information Table” on page 103 for more information). The fields of the
table can be edited and the information therein used for reporting purposes
even if no column identification tags are present. This is often referred to as
soft tagging. When the Column ID Upgrade Kit is installed and a column
identification tag is present, the content of the table corresponds to the
information stored on the tag. This is referred to as hard tagging.
By default, the Column Tag Information table is truncated and only columns up to
and including Injections are displayed. Clicking on >> in the top right corner
displays the whole table.