Agilent InfinityLab LC Series Vialsamplers User Manual 202
7Error Information
Vialsampler Error Messages
Metering home failed
Error ID: 34224
The home position of the metering plunger is not found or the plunger is unable
to move.
Automatic Referencing procedure failed
Error ID: 34225
The execution of the Automatic Referencing procedure was aborted by the
system due to an unexpected error.
Probable cause Suggested actions
1
Blockage in the flow path.
Check if there is a significant pressure difference
(> 10
15 bar) between the mainpass and the
bypass position of the injection valve. If yes, sys
-
tematically replace the components of the main-
pass: Needle seat, Needle, Rotor seal, Sample
loop
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Potential hardware error
Please contact your Agilent service representa-
tive.
2
Physical blockage
Look for potential sources of physical blockage
in the path of the gripper arm and eliminate
them. Restart the sampler and observe if the
problem persists.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals