Agilent InfinityLab LC Series Vialsamplers User Manual 261
8Maintenance
Replace the Peristaltic Pump Cartridge
Replace the Peristaltic Pump Cartridge
When
The tubing is blocked or damaged
The peristaltic cartridge needs to be replaced as part of the yearly maintenance.
Parts required p/n Description
5065-4445 Peristaltic pump with Pharmed tubing
Preparations
Finish any pending acquisition job.
Remove the solvent line from the wash bottle to avoid spilling solvent.
WARNING
When opening capillary or tube fittings solvents may leak out.
The handling of toxic and hazardous solvents and reagents can hold health
risks.
Please observe appropriate safety procedures (for example, goggles,
safety gloves and protective clothing) as described in the material
handling and safety data sheet supplied by the solvent vendor,
especially when toxic or hazardous solvents are used.
NOTE
The peristaltic pump cartridge must be replaced as a single entity. The Pharmed
tubing is not replaceable separately.
1 Open the doors of the module. 2 Press the two clips on the front of the peristaltic pump
cartridge.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals