Agilent InfinityLab LC Series Vialsamplers User Manual 213
7Error Information
Sample Cooler/Sample Thermostat Error Messages
Sample Cooler/Sample Thermostat Error
Messages
Sample temperature control voltage too low, check
fuses and wires
Error ID: 30713
The compressor voltage is below the lower threshold value.
Sample temperature control switched off due to
condensate
Error ID: 30715
The cooler/thermostat was switched off due to a condensate event.
Probable cause Suggested actions
1
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Overfilled container
Empty the condensate container. Verify that the
open end of the tubing doesn’t immerse in the liq
-
uid.
2
Drainage issues
Verify the correct plumbing of the conden-
sate drainage system.
Make sure that no kinks or mechanical
blocks are present in the drainage system.
Avoid the formation of the siphoning effect.
Make sure that the hosting sampler is level.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals