EasyManuals Logo

Agilent Technologies G7129B User Manual

Agilent Technologies G7129B
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #142 background imageLoading...
Page #142 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 142
4 Preparing the Module
Solvent Information
Solvent Information
Observe the following recommendations on the use of solvents.
Follow the recommendations for avoiding the growth of algae, see the pump
manuals.
Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.22
µm filters.
Avoid or minimize the use of solvents that may corrode parts in the flow path.
Consider specifications for the pH range given for different materials such as
flow cells, valve materials etc. and recommendations in subsequent sections.
Recommended Wash Solvents
water
ethanol
methanol
water/acid (especially for basic compounds)
water/base (especially for acidic compounds)
water/acetonitrile
NOTE
For different wash solvents as mentioned above, verify that the wash solvent is
suitable for the silicone wash tubing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals