Agilent InfinityLab LC Series Vialsamplers User Manual 99
3Using the Module
Using the Optional Integrated Column Compartment
Method Parameters
The Method dialog box enables the user to change the method parameters
relevant for the ICC:
• Temperature:
• Not Controlled: The temperature of the column oven stays uncontrolled.
• At specified temperature (°C): The temperature of the column oven is
regulated to stay at the setpoint temperature specified.
• Unchanged: The temperature setting stays unchanged, meaning that the
setting currently stored in the system is used by the method.
•Valve Position/Column: This section provides information about the location
of the column that is being used for the given method.
• Enforce column for run: Marking this checkbox restricts the execution of
the method to a specific column type, identified by its product number.
•Stoptime:
• As Pump/Injector: The ICC uses the same stoptime setting as specified for
the pump or the sampler.
• After specified time (min): The stop time defines a time limit for the
analysis, after which all method parameters are reset to initial values.
•Posttime:
• Off: No post-run period is applied.
• For specified time (min): After the analysis, the system stays in the
post-run state for the specified time. Using a post-time period can be
beneficial when the column needs to be re-equilibrated before starting the
next analysis (for example, gradient elutions).
Valid inputs are 10 to 80 °C in increments of 0.1 °C.
For Stoptime and Posttime, valid inputs are 0.01 to 99999 minutes in increments
of 0.1 min.