Agilent InfinityLab LC Series Vialsamplers User Manual 51
2 Site Requirements and Specifications
Specifications of the Integrated Column Compartment
Specifications of the Integrated Column
Compartment
Physical Specifications Agilent InfinityLab LC Series
Integrated Column Compartment (G7130A)
Table 9 Physical Specifications (G7130A)
Type Specification Comment
Weight 1.8 kg (4.0 lbs)
Dimensions
(height x width x depth)
86.5 x 396.0 x 106.5 mm
(3.4 x 15.6 x 4.2 inches)
Maximum outside
Supply Voltage 24 V DC
Power consumption 110 W
Ambient operating temperature 4 – 55 °C (39 – 131 °F)
Ambient non-operating tem-
perature
-40 – 70 °C (-40 – 158 °F)
Humidity
< 95 % r.h. at 40 °C (104 °F)
1
1
If a sample cooler is included the upper value for humidity can be reduced. Please check your lab condi-
tions to stay beyond dew point values for non–condensing operation.
Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards
IEC, EN, CSA, UL
Installation category II, Pollution
degree 2
For indoor use only
ISM Classification ISM Group 1 Class B According to CISPR 11