Agilent InfinityLab LC Series Vialsamplers User Manual 229
8Maintenance
Exchange the Needle Assembly
Exchange the Needle Assembly
When
• The needle is visibly damaged.
• Leaks or blockages are observed.
• The limit for the needle interaction EMF counter is exceeded.
• The needle needs to be replaced as part of the yearly maintenance.
Tools required p/n Description
8710-1924 Open-end wrench 14 mm
8710-2140 Screwdriver Torx TX-10
Parts required # p/n Description
1 G7129-87200 Needle assembly, 1260 Vialsampler
for G7129-87017 needle seat
OR 1 G7129-87201 Needle assembly, 1290 Vialsampler
for G7129-87012 needle seat
OR 1 G7129-87202 Needle assembly, slotted, for high injection volumes
Preparations
• Finish any pending acquisition job.
• Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent.
• Close the shutoff valves at the pump if available.