Agilent InfinityLab LC Series Vialsamplers User Manual 49
2 Site Requirements and Specifications
Specifications of the Sample Thermostat
Specifications of the Sample Thermostat
The Agilent InfinityLab Sample Thermostat is the combination of an electric
heater and a vapor-compression refrigeration system. It uses isobutane as a
non-Freon refrigerant, which is harmless to the environment and does not affect
the ozone layer and global warming but is combustible. Please adhere to the
warnings listed in the manual.
Table 7 Physical Specifications of the Sample Thermostat
Type Specification Comments
Weight <6 kg (< 13.2 lbs)
Dimensions
(height x width x depth)
205 x 340 x 370 mm
(8.1 x 13.4 x 14.6 inches)
Refrigerant gas R600a (0.030 kg) Ozone depletion potential (ODP) =0
Global warming potential (GWP) =3
Supply voltage 24VDC
Current 10 A max.
Ambient operating
temperature
4 – 40 °C (39104 °F)
Ambient non-operat-
ing temperature
-4070 °C (-40158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards:
IEC, EN, CSA, UL
Installation category II,
Pollution degree 2
For indoor use only
ISM Classification ISM Group 1 Class B According to CISPR 11

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals