EasyManuals Logo

Agilent Technologies G7129B User Manual

Agilent Technologies G7129B
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 159
4 Preparing the Module
Setting up the Vialsampler
Table 16 Control settings
Figure 30 Control settings
The Sampler control parameters are in the following sections:
Missing Vial
Mark the Ignore missing vial check box to specify that, if a
vial is missing, the injector ignores it and continues with a
6-second dummy run. The message "Missing vial <x>" is
logged, and the system continues with the next injection.
Illumination
Toggles the illumination of the sample area, On or Off.
•At Power On
The section is available when a cooler/thermostat is
installed and configured. Mark the Turn on Thermostat
check box to specify that the cooler/thermostat is
switched on automatically when the instrument is
switched on.
•Thermostat
The section is available when a cooler/thermostat is
installed and configured and the Constant temperature
mode is selected.
Select On to switch on the cooler/thermostat. Specify the
required temperature in the adjacent field. The specified
temperature must be at least 5
°C below ambient for
proper temperature control.
Select Off to switch off the cooler/thermostat.
•Automatic Turn On
You can set a date and time at which the cooler/thermo-
stat switches on automatically.
Pump connected to Sampler
Use this section to specify the pump that is used with the
Vialsampler. If more than one pump is configured, display
the drop-down list and select the appropriate pump from
the list.
•Enable Analysis
This feature requires LC & CE Drivers A.02.19 or newer and
is only available for the Sample Thermostat. With this func
-
tion, you can specify if the analyses should start With any
temperature or only when the Temperature is within ± 2 °C
range of the setpoint temperature.
NOTE
The Enable Analysis section is disabled when Not con-
trolled is selected in the Temperature section.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals